Connection "remote"<br />

Inputs<br />

The digital inputs at connection "remote" are used to connect various functions of<br />

the electronic drive unit. Functions are assigned to the inputs DI1 - DI2 ex factory.<br />

These can be configured via interface RS485 and the <strong>Pfeiffer</strong> Vacuum parameter<br />

set.<br />

DI1 (Enable venting) / Pin 2<br />

V+ : Venting is enabled (venting according to venting mode)<br />

open: Venting locked (no venting is performed)<br />

DI Motor pump / Pin 3<br />

After Pin 4 (pumping station) is activated and the electronic drive unit successfully<br />

completes the self-test (duration approx. 8 seconds), the turbopump is placed into<br />

operation. During operation, the turbopump can be switched off and on again,<br />

while the pumping station remains switched on. The turbopump is not vented<br />

thereby.<br />

V+ : Turbopump motor on<br />

open: Turbopump motor off<br />

DI Pumping station / Pin 4<br />

Connected pumping station components (e.g. backing pump, venting valve, air<br />

cooling unit) are triggered and, with Pin 3 (motor) simultaneously activated, the<br />

turbopump is placed in operation. Any ongoing error messages are reset when<br />

their cause has been eliminated.<br />

V+ : Malfunction acknowledgement and pumping station on<br />

open: Pumping station off<br />

DI Standby / Pin 5<br />

In standby mode, the turbopump operates at a specified rotor speed < nominal rotation<br />

speed. Factory setting and recommended operation are 66.7 % of the nominal<br />

rotation speed.<br />

V+ : Standby activated<br />

open: Standby off, operation at nominal rotation speed<br />

DI2 (Heating) / Pin 6<br />

V+ : Heating on<br />

open: Heating off<br />

DI3 (Sealing gas) / Pin 10<br />

V+ : Sealing gas valve open<br />

open: Sealing gas valve closed<br />

DI Error acknowledgement / Pin 13<br />

V+ : Reset ongoing error messages when cause has been eliminated with a pulse<br />

of min. 500 ms duration.<br />

open: Inactive<br />

DI Remote priority / Pin 14<br />

V+ : The connection "remote" has operation priority over all other digital inputs.<br />

open: Remote priority inactive<br />

10