Connection "remote"<br />
Inputs<br />
The digital inputs at connection "remote" are used to connect various functions of<br />
the electronic drive unit. Functions are assigned to the inputs DI1 - DI2 ex factory.<br />
These can be configured via interface RS485 and the <strong>Pfeiffer</strong> Vacuum parameter<br />
set.<br />
DI1 (Enable venting) / Pin 2<br />
V+ : Venting is enabled (venting according to venting mode)<br />
open: Venting locked (no venting is performed)<br />
DI Motor pump / Pin 3<br />
After Pin 4 (pumping station) is activated and the electronic drive unit successfully<br />
completes the self-test (duration approx. 8 seconds), the turbopump is placed into<br />
operation. During operation, the turbopump can be switched off and on again,<br />
while the pumping station remains switched on. The turbopump is not vented<br />
thereby.<br />
V+ : Turbopump motor on<br />
open: Turbopump motor off<br />
DI Pumping station / Pin 4<br />
Connected pumping station components (e.g. backing pump, venting valve, air<br />
cooling unit) are triggered and, with Pin 3 (motor) simultaneously activated, the<br />
turbopump is placed in operation. Any ongoing error messages are reset when<br />
their cause has been eliminated.<br />
V+ : Malfunction acknowledgement and pumping station on<br />
open: Pumping station off<br />
DI Standby / Pin 5<br />
In standby mode, the turbopump operates at a specified rotor speed < nominal rotation<br />
speed. Factory setting and recommended operation are 66.7 % of the nominal<br />
rotation speed.<br />
V+ : Standby activated<br />
open: Standby off, operation at nominal rotation speed<br />
DI2 (Heating) / Pin 6<br />
V+ : Heating on<br />
open: Heating off<br />
DI3 (Sealing gas) / Pin 10<br />
V+ : Sealing gas valve open<br />
open: Sealing gas valve closed<br />
DI Error acknowledgement / Pin 13<br />
V+ : Reset ongoing error messages when cause has been eliminated with a pulse<br />
of min. 500 ms duration.<br />
open: Inactive<br />
DI Remote priority / Pin 14<br />
V+ : The connection "remote" has operation priority over all other digital inputs.<br />
open: Remote priority inactive<br />
10