Product description<br />
3 Product description<br />
3.1 Product identification<br />
Product features<br />
The electronic drive unit <strong>TC</strong> <strong>400</strong> is an integrated component of the turbopump. It's<br />
purpose is to drive, monitor and control the entire pump.<br />
Characteristics <strong>TC</strong> <strong>400</strong><br />
Connection voltage <strong>TC</strong> 24 V DC ± 5 % 48 V DC ± 5 %<br />
Connection panel Standard (RS485) Standard (RS485)<br />
Turbopump HiPace 300, <strong>400</strong>, 700, 800 300, <strong>400</strong>, 700, 800<br />
To correctly identify the product when communicating with <strong>Pfeiffer</strong> Vacuum, always<br />
have the information from the rating plate available.<br />
3.2 Range of application<br />
<strong>Pfeiffer</strong> Vacuum electronic drive units <strong>TC</strong> <strong>400</strong> must be installed and operated in the<br />
following ambient conditions.<br />
Installation location<br />
weather protected (indoors)<br />
Protection category IP 54<br />
Protection class<br />
III<br />
Temperature<br />
+5 °C to +40 °C (up to +35 °C with air cooling)<br />
Relative humidity max. 80 %, at T ≤ 31 °C, up to max. 50% at T ≤ 40 °C<br />
Atmospheric pressure:<br />
77 kPa - 106 kPa<br />
Installation altitude<br />
2000 m max.<br />
Degree of pollution 2<br />
Overvoltage category<br />
II<br />
3.3 Function<br />
b<br />
d<br />
e<br />
a<br />
b<br />
c<br />
Fig. 1: Standard panel for the <strong>TC</strong> <strong>400</strong><br />
a Mains connection "DC in"<br />
b Connection "accessory A+B"<br />
c Connection "remote"<br />
d<br />
e<br />
Service connection "PV.can"<br />
Connection "RS485"<br />
6