The <strong>Pfeiffer</strong> Vacuum parameter set<br />
Parameter [P:017] = 1<br />
When the pumping station [P:010] is switched on, the rotation speed switch point 1<br />
is the signal generator. When the pumping station is switched off, signal output<br />
and status query are based on the rotation speed switch point 2. The signal output<br />
is governed by the hysteresis between the two switch points.<br />
f (%)<br />
[P:017] = 1<br />
[P:701]<br />
[P:719]<br />
t<br />
[P:010]<br />
1<br />
0<br />
t<br />
[P:302]<br />
1<br />
0<br />
t<br />
Process<br />
Fig. 6: Example for the configuration rotation speed switch point 1+2 active; [P:701] > [P:719]<br />
f (%)<br />
[P:017] = 1<br />
[P:719]<br />
[P:701]<br />
t<br />
1<br />
[P:010]<br />
0<br />
t<br />
1<br />
[P:302]<br />
t<br />
0<br />
Process<br />
Fig. 7: Example for the configuration rotation speed switch points 1+2 active; [P:701] < [P:719]<br />
Rotation speed setting<br />
mode<br />
The rotation speed setting mode reduces the rotation speed and hence the<br />
throughput of the turbopump. The pumping speed of the turbopump changes proportional<br />
to rotation speed. Standby mode is ineffective during rotation speed setting<br />
mode. The set rotation speed is adjusted by the set value in rotation speed setting<br />
mode [P:707]. The rotation speed switch point varies with the set rotation<br />
speed. Underrunning or overrunning the set value in rotation speed setting mode<br />
activates and deactivates the status signal [P:306] SetSpdAtt respectively.<br />
Adjust the parameter [P:707] to the desired value in %.<br />
Parameter [P:026] = 1<br />
Read the parameters [P:308]/[P:397].<br />
Standby<br />
<strong>Pfeiffer</strong> Vacuum recommends standby mode for the turbopump during process<br />
and production stops. When standby mode is active, the electronic drive unit reduces<br />
the rotation speed of the turbopump. The factory setting for the set value in<br />
standby mode is 66.7 % of the nominal rotation speed. Underrunning or overrunning<br />
the set speed in standby mode activates or deactivates the status signal<br />
[P:306] SetSpdAtt.<br />
Adjust the parameter [P:717] to the desired value in %.<br />
Parameter [P:002] = 1<br />
Read the parameters [P:308]/[P:397].<br />
22