The <strong>Pfeiffer</strong> Vacuum parameter set<br />

Parameter [P:017] = 1<br />

When the pumping station [P:010] is switched on, the rotation speed switch point 1<br />

is the signal generator. When the pumping station is switched off, signal output<br />

and status query are based on the rotation speed switch point 2. The signal output<br />

is governed by the hysteresis between the two switch points.<br />

f (%)<br />

[P:017] = 1<br />

[P:701]<br />

[P:719]<br />

t<br />

[P:010]<br />

1<br />

0<br />

t<br />

[P:302]<br />

1<br />

0<br />

t<br />

Process<br />

Fig. 6: Example for the configuration rotation speed switch point 1+2 active; [P:701] > [P:719]<br />

f (%)<br />

[P:017] = 1<br />

[P:719]<br />

[P:701]<br />

t<br />

1<br />

[P:010]<br />

0<br />

t<br />

1<br />

[P:302]<br />

t<br />

0<br />

Process<br />

Fig. 7: Example for the configuration rotation speed switch points 1+2 active; [P:701] < [P:719]<br />

Rotation speed setting<br />

mode<br />

The rotation speed setting mode reduces the rotation speed and hence the<br />

throughput of the turbopump. The pumping speed of the turbopump changes proportional<br />

to rotation speed. Standby mode is ineffective during rotation speed setting<br />

mode. The set rotation speed is adjusted by the set value in rotation speed setting<br />

mode [P:707]. The rotation speed switch point varies with the set rotation<br />

speed. Underrunning or overrunning the set value in rotation speed setting mode<br />

activates and deactivates the status signal [P:306] SetSpdAtt respectively.<br />

Adjust the parameter [P:707] to the desired value in %.<br />

Parameter [P:026] = 1<br />

Read the parameters [P:308]/[P:397].<br />

Standby<br />

<strong>Pfeiffer</strong> Vacuum recommends standby mode for the turbopump during process<br />

and production stops. When standby mode is active, the electronic drive unit reduces<br />

the rotation speed of the turbopump. The factory setting for the set value in<br />

standby mode is 66.7 % of the nominal rotation speed. Underrunning or overrunning<br />

the set speed in standby mode activates or deactivates the status signal<br />

[P:306] SetSpdAtt.<br />

Adjust the parameter [P:717] to the desired value in %.<br />

Parameter [P:002] = 1<br />

Read the parameters [P:308]/[P:397].<br />