Connection "RS485"<br />

6 Connection "RS485"<br />

6.1 Connections<br />

A <strong>Pfeiffer</strong> Vacuum display and control panel (DCU or HPU) or an external PC can be<br />

connected to the electronic drive unit via the connection designated "RS485". The<br />

interface is electrically isolated from the maximum supply voltage of the electronic<br />

drive unit. The electrical couplings are optically decoupled internally.<br />

Désignation<br />

Valeur<br />

Interface série<br />

RS485<br />

Vitesse de transmission<br />

9600 bauds<br />

Longueur d'un mot de données 8 bits<br />

Parité<br />

aucune (no parity)<br />

Bits de départ 1<br />

Bits d'arrêt 1..2<br />

5 Pin Assignment<br />

1 2 1 RS485: D+<br />

2 +24 V output, loadable with ≤ 210 mA (with remote - pin 1)<br />

3 GND<br />

4 RS485: D-<br />

4 3<br />

5 not connected<br />

6.2 Connecting <strong>Pfeiffer</strong> Vacuum display and control units or PC<br />

Use the connection cable supplied with the control panel or from the range of<br />

accessories.<br />

The connection of respectively one external operating unit is possible on the interface<br />

RS 485.<br />

A USB interface (PC) can be connected via the USB/RS485-converter.<br />

USB/RS485-converter<br />

DCU 310 / <strong>400</strong><br />

<strong>TC</strong> <strong>400</strong> / TM 700<br />

DCU 002<br />

RS485<br />

HPU 001<br />

Fig. 3:<br />

Optional connection possibilities for interface RS485<br />