<strong>Pfeiffer</strong> Vacuum Protocol for "RS485"<br />
Example 2<br />
Control command<br />
Switch on pumping station (parameter [P:010], device address slave: "042")<br />
◦! 0 4 2 1 0 0 1 0 0 6 1 1 1 1 1 1 0 2 0 C<br />
R<br />
ASCII 48 52 50 49 48 48 49 48 48 54 49 49 49 49 49 49 48 50 48 13<br />
Control command understood<br />
Switch on pumping station (parameter [P:010], device address slave: "042")<br />
◦! 0 4 2 1 0 0 1 0 0 6 1 1 1 1 1 1 0 2 0<br />
C<br />
R<br />
ASCII 48 52 50 49 48 48 49 48 48 54 49 49 49 49 49 49 48 50 48 13<br />
8.3 Applied data types<br />
Data type Description Size l1 - l0 Example<br />
0 False / true 06 000000 / 111111<br />
1 Positive integer number 06 000000 to 999999<br />
2 Positive fixed comma number 06 001571 equal to 15,71<br />
4 Symbol chain 06 <strong>TC</strong>_<strong>400</strong><br />
7 Positive integer number 03 000 to 999<br />
11 Symbol chain 16 BrezelBier&Wurst<br />