HT Glycan Assay Per<strong>for</strong>mance Specifications<br />
8ul with concentration range <strong>of</strong> 1.25mg/ml to<br />
Amount <strong>of</strong> sample required:<br />
7.5mg/ml (10ug to 60ug <strong>of</strong> MAb total).<br />
Reproducibility <strong>of</strong> %area:<br />
CV < 10% <strong>for</strong> a peak >= 2.5% <strong>of</strong> total glycan.<br />
1ng <strong>of</strong> G0F standard. (Smallest amount <strong>of</strong> labeled<br />
Limit <strong>of</strong> detection:<br />
G0f standard that can be detected.)<br />
>95% <strong>of</strong> all N-Linked glycans will be released from<br />
Deglycosylation:<br />
mAb.<br />
Appropriate <strong>for</strong> neutral glycans found on MAbs.<br />
Usable size range: Some charged glycans may run outside <strong>of</strong> our<br />
usable range.<br />
Sizing ing reproducibility: CV < 2.5%<br />
< 0.5% <strong>of</strong> previous sample can be detected in<br />
Carryover:<br />
following sample.<br />
Sample prep, chip prep, and < 8 hours <strong>for</strong> a 96-well plate.<br />
analysis time:<br />
Number <strong>of</strong> samples per chip Up to 192 samples be<strong>for</strong>e chip needs to be<br />
prep: reloaded with fresh gel.<br />
Number <strong>of</strong> samples per chip Up to 384 samples be<strong>for</strong>e chip needs to be<br />
lifetime: replaced.<br />