15<br />

Model TLV25<br />

REF 00T0650 (230V/500-60Hz), REF 00T0780 (115V/60Hz)<br />

MAIN CHARACTERISTICS<br />

The TLV25 system contains a 25 litres Dewar flask. The fluid in the flask must permanently be cooled by a<br />

separate cryostat. The temperature set point is maintained via a microprocessor controlled heating<br />

element. When using the KV40 or KV80 minimum working temperatures of minus 40°C or minus 80°C can<br />

be reached. All presented data is measured by using a TLV25 filled with 25 litres of methanol and a KV40<br />

or KV80 immersion cooler. At the minimum-temperature, the provided heat removal is still enough to<br />

maintain stable temperature control, even when glasswork is placed in the bath. The systems accuracy is<br />

better than the requirements of ASTM D 445 and ISO 3104. The bath is illuminated by a fluorescent light<br />

built in behind the Dewar-flask. The top lid has a turn table construction with 3 holes + 3 lids of 51 Ø<br />

mm. By turning the lid the immersed viscometer can be positioned in front of the window. This window is<br />

heated to keep clear sight at low temperatures. The set point can be set in steps of 0.1°C from -90°C up<br />

to +60°C (-130..140°F). Overall accuracy is better than ±0.03K.<br />

Range<br />

-90..60°C/-130..140°F<br />

Reading<br />

Standard °C, °F on<br />

request<br />

Used materials inside<br />

bath<br />

Stainless Steel 304,<br />

PTFE<br />

Setting ±<br />

0.1°C<br />

Offset<br />

0.01°C<br />

Stability ± 0.04K<br />

Heating<br />

780 W<br />

Heaters 1<br />

Bath volume<br />

25 L<br />

Number of lids 3<br />

Opening lid<br />

51 (dia) mm<br />

Opening bath<br />

260 * 240 mm<br />

Depth bath<br />

400 mm<br />

TLV25<br />

OPTIONS<br />

REF 00T0212 KV40 to reach -40°C (230V/50Hz), REF 00T212A (230V/60Hz), REF 00T0259 (115V/60Hz)<br />

REF 00T0216 KV80 to reach -80°C (230V/50Hz), REF 00T216A (230V/60Hz), REF 00T0260 (115V/60Hz)<br />

ACCESSORIES<br />

REF 10T6065 Viscohanger<br />

REF 10T6001 Viscometer holder ‘Ubbelohde’<br />

REF 10T6061 Viscometer holder ‘Cannon Fenske Routine’<br />

REF 10T6071 Viscometer holder ‘Cannon Fenske Reversed Flow’<br />

REF 10T6321 Viscometer holder ‘Utube (Ostwald) Routine’<br />

REF 10T6051 Viscometer holder ‘Utube (Ostwald) Reversed Flow’<br />

REF 10T6040 Viscometer holder ‘Pinkevitch transparent 269mm<br />

REF 10T6050 Viscometer holder ‘Pinkevitch transparent long’<br />

REF 10T6090 8 Channel Stop-watch<br />

REF 00T0239 ASTM Thermometer holder<br />

Viscometer/glass capillaries, see special section ‘Viscometers’<br />

ASTM Thermometers, see special section ‘ASTM Thermometers’<br />

Viscosity Reference oil, see special section ‘Viscosity Calibration Standards’<br />

KV40 & KV80<br />

MAIN SPARE PARTS 230V/50-60Hz (115V/60Hz ON REQUEST)<br />

REF 04T0100 Pump assembly with motor<br />

REF 25T0210 Heater 1500 W<br />

REF 28T3015 Temperature controller 85..250V<br />

REF 25T2320 PT-100 115 mm<br />

REF 24T9250 Fluorescent lamp<br />

REF 06T0377 PCB window heating<br />

REF 25T0140 Window heating 115V 40W (2 placed in series 230V)<br />

REF 28T0091 Sensor programmed for window heating<br />

REF 06T1870 PCB filter temp controller<br />

REF 24T8420 Solid state<br />

REF 24T8585 Thermostat<br />

REF 26T1420 Dewar flask<br />

<strong>Tamson</strong> Instruments B.V. T. + 31 10 522 43 73 VAT: NL 80 66 34 984 B01<br />

Van ’t Hoffstraat 12 F. + 31 10 521 19 41 Rabobank: 16 01 00 046<br />

2665 JL BLEISWIJK www.tamson.com Chamber of Commerce: 271 69 541<br />

The Netherlands<br />

sales@tamson.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!