25<br />

THERMOMETER<br />

<strong>Tamson</strong> Thermometer 3 decimals (TT3)<br />

IEC 5293<br />

APPLICATION<br />

The calibration of probes, thermometers, instruments, and indicators and temperature measurement of<br />

viscosity baths.<br />

MAIN CHARACTERISTICS<br />

The <strong>Tamson</strong> Thermometer has a three decimal reading. The thermometer uses a standard PT100 probe for<br />

measurements from -20 ..+250°C or -4 ..+482°F. The thermometer uses an internal correction table to meet<br />

the IEC751 required accuracy of ± 0.01°C. A high accuracy data converter samples the PT100 value and a<br />

microprocessor is used to correct these values. The calibration data is kept in internal memory. A second<br />

memory offers the possibility for data logging. Sampling times can vary from 1 second to 300 seconds. At each<br />

interval the temperature is measured and stored. Each measurement has an individual time stamp and can be<br />

recalled easily using the PC software. The thermometer operates on a Battery pack. A full charge supplies<br />

power for 16 days of continuous use. When off the battery will keep its charge.<br />

Item Unit TT3<br />

Ordering code<br />

10T6095<br />

Range<br />

Reading<br />

Interface<br />

-100..+300°C/-148..572°F<br />

°C or °F menu selectable<br />

RS232<br />

Resolution [°C/F] 0.001<br />

Accuracy [°K] ± 0.01<br />

Linearity [°K] ± 0.01<br />

Power<br />

Charge<br />

Mains adapter<br />

Sleep current<br />

Nominal<br />

current<br />

Battery pack<br />

Battery NiMH operated<br />

2Hrs<br />

100-240~50/60Hz 9V-1A<br />

0.01mA<br />

2mA<br />

800mAh<br />

Log time [hr] 19hr sample rate 1s<br />

5700hr sample rate 300s<br />

Sample rate [sec] 1,2,5,10,20,60,120,300<br />

Dimensions [mm] 105x32x150 (LxWxH)<br />

excl probe<br />

Weight [gr] 355 (inclusive probe)<br />

Power [Watt] Battery NiMH operated<br />

Frequency [Hz] Suited for both 50 & 60<br />

CE<br />

All models conform to CE regulation<br />

TT-3<br />

TT-3 + Suitcase<br />

Necessary Accessories<br />

REF 25T2350 Temperature probe* -100..300°C/-148..302°F 115 x 3mm, Class "a"<br />

REF 21T0200 Blue Suitcase to store TT-3.<br />

* The unit will be calibrated in combination with the supplied probe in order to meet IEC 751. It is not possible to exchange<br />

probes.<br />

Options<br />

REF 10T6096<br />

REF 10T6097<br />

REF 10T6098<br />

REF 24T0038<br />

REF 24T0037<br />

REF 24T0036<br />

Works certificate 0 .. +150°C<br />

Works certificate -40 .. +150°C<br />

Verification calibration per point [-40 .. +150°C]<br />

RS232 .. USB converter<br />

Flatcable 2mtr<br />

Flatcable 1mtr<br />

<strong>Tamson</strong> Instruments B.V. T. + 31 10 522 43 73 VAT: NL 80 66 34 984 B01<br />

Van ’t Hoffstraat 12 F. + 31 10 521 19 41 Rabobank: 16 01 00 046<br />

2665 JL BLEISWIJK www.tamson.com Chamber of Commerce: 271 69 541<br />

The Netherlands<br />

sales@tamson.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!