<strong>ZSX</strong>100e <strong>ZSX</strong>101e <strong>ZSX</strong>100s <strong>ZSX</strong>101s<br />

Area limiting diaphragm<br />

Divergence slit<br />

30 mm N fixed (no attenuator) or 35, 30, 25, 20, 10,<br />

3, 1, 0.5 mm N and attenuator exchanger<br />

(For spinner or r-2 driving)<br />

Standard resolution, 1 kind, fixed or slit exchanger<br />

Select from standard resolution, high resolution and<br />

ultralight element use<br />

30 mm N fixed (no attenuator) or 35, 30, 25, 20, 10<br />

mm N (no attenuator exchanger)<br />

(spinner)<br />

Standard resolution, 1 kind, fixed or slit exchanger<br />

Selection from standard resolution and high<br />

resolution<br />

Receiving slit<br />

Goniometer<br />

Crystal exchanger<br />

Analyzing crystal<br />

Detector<br />

Vacuum system<br />

Temperature stabilizer<br />

Option<br />

High resolution for SC (standard) high sensitivity for SC (optional), for F-PC (select for <strong>ZSX</strong>101e/101s)<br />

2-22-22 3-axis independent drive system<br />

Scan range: SC -1~118E (22)<br />

F-PC 7~148E (22)<br />

Scan speed: Max 1400E/min<br />

Step angle: 1E/1000, 2E/1000, 5E/1000, 1E/100, 2E/100, 5E/100, 1E/100<br />

Continuous scan: 0.1~240E /min (optional)<br />

10-crystal exchanger or 4-crystal exchanger<br />

Select from LIF(200) PET, Ge, TAP, RX-4, RX-9, RX-35, LiF (220)<br />

Select from RX-40, RX-45, RX-60, RX-70, RX-80<br />

For Ti~U: Scintillation counter (SC)<br />

For Be~Zn: Gas-flow proportional counter (F-PC)<br />

With center wire cleaner and gas density stabilizer<br />

High-speed vacuum system (2 vacuum pumps), or general-purpose vacuum system (1 vacuum pump)<br />

Temperature for control: 36.5EC"0.1EC<br />

Powder sample attachment, automatic He substitution device, PR gas switching device, remote diagnosis,<br />

remote control, mapping device, sample observation mechanism<br />

Vol. 16 No. 2 1999 63