CDMA Over Air<br />

CDMA Over Air<br />

Optional CDMA Over Air <strong>Measurement</strong> Settings<br />

Setting Average, Sweep, and Restart<br />

Setting the Sweep<br />

You can select either continuous or single sweep. Continuous sweep provides<br />

repeated, ongoing measurements. Single sweep performs a single measurement that<br />

is repeated only when you push the Single button.<br />

To set sweep:<br />

Step<br />

Notes<br />

1 [Average/Sweep]<br />

2 [Continuous] or [Single] Each time you press this softkey, the<br />

selected option changes.<br />

Setting Averaging<br />

You can choose to have averaging on or off and set the number of averages you want<br />

to take. Averaging only applies to the data, not the trace. You can select from the<br />

following types of averaging: When initialized, the <strong>Agilent</strong> <strong>E7495A</strong>/B defaults to<br />

running average:<br />

• A running average computes the new result as the weighted sum of the last result<br />

and the new measurement. The last result is weighted by (n – 1)/n. The new<br />

result is weighted by 1/n. Each new measurement produces a new result.<br />

To set averaging:<br />

Step<br />

Notes<br />

1 Set the number of averages.<br />

a. [Average/Sweep]<br />

b. [Average]<br />

c. Enter the number of averages<br />

using the numeric keypad.<br />

d. [Enter]<br />

234 Chapter 8