14<br />

Warning:<br />

Do not place the blades axis on some hard surface, as this might<br />

seriously damage the blades!<br />

- Put the mesher (bridge opened) in the sterilization container.<br />

- All parts can be sterilized with saturated steam at the common<br />

temperature of 134 o C (273 o F), usually for 5-10 minutes, or at 121 o C<br />

(250 o F) for prolonged exposure time of approx. 20 minutes.<br />

II-4. Lubrication<br />

For long service life of the instrument, it is recommended to put some<br />

drops of surgical oil into the two holes (10) on the top of the bridge (next to<br />

the grip) after washing <strong>and</strong> before sterilization. <strong>Humeca</strong> supplies<br />

STERILIT ® oil for that purpose.<br />

II-5. First use<br />

II-5.1. Check carrier adjustment<br />

The mesher is suitable for different kinds of skin graft <strong>carriers</strong>.<br />

It was adjusted by <strong>Humeca</strong> on request to fit V10- or V15-<strong>carriers</strong>.<br />

If the instrument was adjusted for <strong>Humeca</strong> V10 <strong>carriers</strong>, it is also<br />

compatible with Zimmer ® <strong>carriers</strong>.<br />

If it was adjusted to <strong>Humeca</strong> V15 <strong>carriers</strong> it is also compatible with<br />

Aesculap ® / B.Braun ® <strong>carriers</strong>.<br />

To check the adjustment of the mesher, please proceed as follows:<br />

- Put the mesher on a table with the backside towards you, put your h<strong>and</strong><br />

under the transport axis <strong>and</strong> lift it a little, as shown in fig. 11.<br />

- If you can lift the axis a little (only about 0.5 mm or 0.02”) <strong>and</strong> so the<br />

axis has some play, the mesher is adjusted to <strong>Humeca</strong> ® V-15 <strong>and</strong><br />

Aesculap ® / B.Braun ® <strong>carriers</strong>.<br />

- If there’s no vertical movement in the axis, the mesher is adjusted to<br />

<strong>Humeca</strong> ® V-10 <strong>and</strong> Zimmer ® <strong>carriers</strong>.