Please follow the procedure described below for meshing:<br />
- Place the graft on the carrier (preferred is epidermal side down) <strong>and</strong> put<br />
it in the mesher in the direction indicated by the ‘arrow’ of the grooves.<br />
- Push the carrier in the direction of the blades, while at the same time<br />
turning the h<strong>and</strong>le clockwise until the carrier moves forward.<br />
- Now gently keep turning the h<strong>and</strong>le to move the carrier through the<br />
mesher.<br />
- After meshing, use a forceps to remove the graft from the carrier.<br />
- After complete pass, remove the carrier from the mesher <strong>and</strong> apply the<br />
graft to the wound.<br />
NOTE (1):<br />
If during meshing the graft moves with the blades of the mesher, please<br />
use a forceps to remove it from the blades <strong>and</strong> place it on the carrier<br />
again. Be careful not to damage the blades.<br />
NOTE (2):<br />
If the graft length exceeds the length of the carrier, first mesh part of it<br />
on a V-carrier <strong>and</strong> then continue meshing by connecting a second<br />
carrier to the first one. The grooves connect to each other <strong>and</strong> if care is<br />
taken (seam between both <strong>carriers</strong> minimal), the mesh pattern will not<br />
be interrupted.<br />
Another well known method of meshing a long stripe of graft is to fold it<br />
on the carrier double down <strong>and</strong> mesh as usual.<br />
If more grafts have to be meshed, the same carrier may be used several<br />
times. However do not re-sterilize it <strong>and</strong> use it again at another surgery.<br />
Quality <strong>and</strong> cleanness of the product cannot be guaranteed after resterilization.<br />
I-6. Labelling<br />
Fig. 7 explains the meaning of the signs on the label on the package of a<br />
<strong>Humeca</strong> V-carrier. It is important to take notice of the expiry date <strong>and</strong> LOT<br />
number. In case of any complaint, you will have to refer to these data.<br />
Sterility cannot be guaranteed when the package is damaged. Please do<br />
not use the product in such case.