All new calibration values will automatically override the existing data. It is<br />
recommended to calibrate the meter periodically and or if it is suspected to be<br />
inaccurate.<br />
Always rinse the probe with either DI water or rinse solution before and after<br />
each calibration/sample measurement. When calibrating in air, make sure that<br />
any water droplets from the probe’s membrane are removed.<br />
3.4 Accessing Calibration mode<br />
From measurement mode, press CAL (F2) key. The meter goes to<br />
corresponding calibration mode, based on the selected measurement mode. If<br />
the meter is password protected, you will be prompted to enter password.<br />
3.4.1 Accessing Calibration mode when password protection enabled<br />
1. Make sure you are in measurement mode. If required, press MODE (F3) to<br />
switch to the measurement mode for which you wish to perform calibration.<br />
2. Press CAL (F2) to go to calibration mode. Login Password screen appears<br />
(Figure 9). The meter expects the 5-digit password specified in system<br />
setup.<br />
Figure 9 : Login Password<br />
Note: You can enter ‘00000’ (read-only password) if you wish to view the<br />
calibration report of the last calibration. You are not allowed to perform<br />
calibration when you enter ‘read-only password’.<br />
1. Press up & down arrow keys to enter the first digit of the password and<br />
then press NEXT (F3) key to move to the next digit.<br />
2. The next digit is selected. Press up & down arrow keys to enter the<br />
second digit of the password.<br />
3. Similarly enter all 5-digits.<br />
4. Press ENTER key to confirm the password.<br />
5. When the correct password is entered, the ‘Calibration – Rinse Electrode’<br />
screen will appear.<br />
Note: If you enter an incorrect password, the screen shows ‘Try again’. If<br />
an incorrect password is entered for 3 consecutive times, the meter goes<br />
to measurement mode<br />
19