27<br />
Situation Visual LED indication Audible signals <strong>Use</strong>r action<br />
In the ACTIVE (continuous) mode<br />
In the ACTIVE (30:2) mode<br />
Too small a patient. Trying to enter<br />
the PAUSE mode or any of the AC-<br />
TIVE modes when the Suction Cup is<br />
in a lower position than <strong>for</strong> the minimum<br />
patient size (sternum height<br />
approximately 17 cm)<br />
The ACTIVE (continuous)<br />
key, LUCAS per<strong>for</strong>ms<br />
continuous chest compressions.<br />
The green<br />
LED signal will blink 8<br />
times per minute<br />
The ACTIVE (30:2) LED<br />
shows an intermittant<br />
green light during compressions<br />
number 26,<br />
27, 28, 29 and 30.<br />
None<br />
Alarm signal alert during<br />
compressions<br />
number 28 (“ding”), 29<br />
(“ding”) and 30<br />
(“dong”).<br />
This is to indicate<br />
when to ventilate<br />
during going compressions.<br />
This is to alert the<br />
operator to ventilate<br />
the patient when<br />
LUCAS temporarily<br />
stops the compressions<br />
at number 30.<br />
None 3 fast signals Continue with<br />
manual compressions.<br />
8.2 Battery replacement and Smart Restart feature<br />
If you change the Battery quickly in 60 seconds or less, with LUCAS in the ON mode, the LUCAS<br />
Smart Restart feature remembers the settings and Start Position according to the table below. If<br />
the Battery change takes more than 60 seconds, LUCAS does a self test and you must adjust the<br />
Start Position again.<br />
Mode when you remove the Battery<br />
PAUSE<br />
Mode when the new Battery is in place again<br />
PAUSE (with the same Start Position)<br />
ACTIVE (continuous)<br />
PAUSE (with the same Start Position)<br />
ACTIVE (30:2)<br />
PAUSE (with the same Start Position)<br />
ADJUST<br />
ADJUST<br />
OFF<br />
OFF<br />
LUCAS TM 2 Chest Compression System – <strong>Instructions</strong> <strong>for</strong> <strong>Use</strong><br />
100666-01 A, Valid from CO J2403, © 2009 JOLIFE AB