Situation Visual LED indication Audible signals <strong>Use</strong>r action<br />

In the ACTIVE (continuous) mode<br />

In the ACTIVE (30:2) mode<br />

Too small a patient. Trying to enter<br />

the PAUSE mode or any of the AC-<br />

TIVE modes when the Suction Cup is<br />

in a lower position than <strong>for</strong> the minimum<br />

patient size (sternum height<br />

approximately 17 cm)<br />

The ACTIVE (continuous)<br />

key, LUCAS per<strong>for</strong>ms<br />

continuous chest compressions.<br />

The green<br />

LED signal will blink 8<br />

times per minute<br />

The ACTIVE (30:2) LED<br />

shows an intermittant<br />

green light during compressions<br />

number 26,<br />

27, 28, 29 and 30.<br />

None<br />

Alarm signal alert during<br />

compressions<br />

number 28 (“ding”), 29<br />

(“ding”) and 30<br />

(“dong”).<br />

This is to indicate<br />

when to ventilate<br />

during going compressions.<br />

This is to alert the<br />

operator to ventilate<br />

the patient when<br />

LUCAS temporarily<br />

stops the compressions<br />

at number 30.<br />

None 3 fast signals Continue with<br />

manual compressions.<br />

8.2 Battery replacement and Smart Restart feature<br />

If you change the Battery quickly in 60 seconds or less, with LUCAS in the ON mode, the LUCAS<br />

Smart Restart feature remembers the settings and Start Position according to the table below. If<br />

the Battery change takes more than 60 seconds, LUCAS does a self test and you must adjust the<br />

Start Position again.<br />

Mode when you remove the Battery<br />

PAUSE<br />

Mode when the new Battery is in place again<br />

PAUSE (with the same Start Position)<br />

ACTIVE (continuous)<br />

PAUSE (with the same Start Position)<br />

ACTIVE (30:2)<br />

PAUSE (with the same Start Position)<br />

ADJUST<br />

ADJUST<br />

OFF<br />

OFF<br />

LUCAS TM 2 Chest Compression System – <strong>Instructions</strong> <strong>for</strong> <strong>Use</strong><br />

100666-01 A, Valid from CO J2403, © 2009 JOLIFE AB