Power Indicator<br />
<strong>The</strong> Power indicator is located on the left hand side of the front panel and<br />
will glow blue when the power is connected properly. If the light does not<br />
illuminate when the <strong>Gas</strong> <strong>Analyzer</strong> is switched on then the power cable may<br />
not be connected properly or a fuse may have blown. If the internal fuses<br />
are suspected of having blown, the unit should be returned to your local<br />
<strong>ADInstruments</strong> representative.<br />
Status Indicator<br />
<strong>The</strong> Status indicator glows green when the <strong>Gas</strong> <strong>Analyzer</strong> is connected to a<br />
PowerLab and has been recognized by LabChart.<br />
Pump On Indicator and Switch<br />
<strong>The</strong> pump is activated using the switch. <strong>The</strong> indicator glows yellow, and the<br />
internal pump may be heard, when it is running.<br />
<strong>Gas</strong> Sample Inlet<br />
A sample inlet port is provided on the front panel. <strong>The</strong> inlet port is directly<br />
connected to the internal gas transducers of the <strong>Gas</strong> <strong>Analyzer</strong>. A sample flows<br />
through this port to be analyzed. Note that the hydrophobic filter, supplied<br />
with every unit, must be used at all times<br />
Flow Control<br />
A single-turn knob on the front panel controls the sampling flow rate to suit<br />
different applications.<br />
<strong>The</strong> Back Panel<br />
Figure 1–2<br />
<strong>The</strong> back panel of the<br />
<strong>Gas</strong> <strong>Analyzer</strong><br />
<strong>Gas</strong> sample<br />
exhaust port<br />
O 2 analog<br />
output<br />
CO 2 analog<br />
output<br />
External<br />
earth<br />
Power socket<br />
and switch<br />
I 2 C connection from the<br />
PowerLab or previous front-end<br />
I 2 C connection to a<br />
further front-end<br />
Cooling fan<br />
outlet<br />
Rating<br />
information<br />
Chapter 1 Overview<br />
15