Power indicator fails to light<br />

<strong>The</strong> <strong>Gas</strong> <strong>Analyzer</strong> is off or the power is switched off at the wall, the power<br />

cable is not connected firmly, or a fuse is blown.<br />

••<br />

Physically check switches and power connections.<br />

••<br />

Check that the power cable is firmly connected at the back of the <strong>Gas</strong><br />

<strong>Analyzer</strong>.<br />

If the internal fuses are suspected of having blown, the unit should be<br />

returned to your <strong>ADInstruments</strong> distributor for servicing.<br />

Drift in the Output Signal<br />

Positive Drift: Contamination build up in the sample chambers of the<br />

transducers will cause a positive drift proportional to the build up. Return<br />

the unit for cleaning or transducer replacement.<br />

Cyclic Drift: Cyclic drift is directly related to temperature variations.<br />

No or Noisy Output<br />

Erratic or noisy output: Excessive bench temperature fluctuations lead to<br />

noisy or erratic drift in the transducers. Place a shroud over the benchtop to<br />

reduce fluctuations. Check power supply for ripples.<br />

No output response or no output at situation: <strong>The</strong> sample cell is filled with<br />

condensate or there is a malfunction of the detector. Return the unit for<br />

cleaning or transducer replacement.<br />

Output extremely erratic: Operating temperature outside specification.<br />

Impeded sample airflow<br />

<strong>The</strong> bacterial filter may become clogged under high humidity conditions.<br />

Replace the bacterial filter.<br />

<strong>The</strong> exhaust port may be blocked or airflow impeded. Ensure the port is not<br />

restricted and is clear of obstructions.<br />

No response to changes in sample concentration<br />

Check that the pump is turned on.<br />

28 <strong>Gas</strong> <strong>Analyzer</strong> Owner’s Guide