Power indicator fails to light<br />
<strong>The</strong> <strong>Gas</strong> <strong>Analyzer</strong> is off or the power is switched off at the wall, the power<br />
cable is not connected firmly, or a fuse is blown.<br />
••<br />
Physically check switches and power connections.<br />
••<br />
Check that the power cable is firmly connected at the back of the <strong>Gas</strong><br />
<strong>Analyzer</strong>.<br />
If the internal fuses are suspected of having blown, the unit should be<br />
returned to your <strong>ADInstruments</strong> distributor for servicing.<br />
Drift in the Output Signal<br />
Positive Drift: Contamination build up in the sample chambers of the<br />
transducers will cause a positive drift proportional to the build up. Return<br />
the unit for cleaning or transducer replacement.<br />
Cyclic Drift: Cyclic drift is directly related to temperature variations.<br />
No or Noisy Output<br />
Erratic or noisy output: Excessive bench temperature fluctuations lead to<br />
noisy or erratic drift in the transducers. Place a shroud over the benchtop to<br />
reduce fluctuations. Check power supply for ripples.<br />
No output response or no output at situation: <strong>The</strong> sample cell is filled with<br />
condensate or there is a malfunction of the detector. Return the unit for<br />
cleaning or transducer replacement.<br />
Output extremely erratic: Operating temperature outside specification.<br />
Impeded sample airflow<br />
<strong>The</strong> bacterial filter may become clogged under high humidity conditions.<br />
Replace the bacterial filter.<br />
<strong>The</strong> exhaust port may be blocked or airflow impeded. Ensure the port is not<br />
restricted and is clear of obstructions.<br />
No response to changes in sample concentration<br />
Check that the pump is turned on.<br />