2. Fill the Capillary Tube When the flashing test strip<br />

appears on the meter’s<br />

display, stick the fingertip.<br />

Then touch the capillary tube<br />

to the blood drop. Keep the<br />

tube level, and allow it to fill<br />

by capillary action. Fill the<br />

tube only halfway.<br />

Avoid getting air bubbles into<br />

the sample. Do not touch the<br />

hole at the top of the bulb. If<br />

blood gets into the bulb during<br />

sample collection, discard the<br />

bulb and repeat the test with a<br />

new test strip and fingerstick.<br />

3. Dose the Test Strip Within 15 seconds of<br />

sticking the fingertip, apply<br />

the blood to the test strip. Put<br />

your finger over the hole at<br />

the end of the capillary tube.<br />

Hold the capillary tube directly<br />

over the target area of the<br />

test strip. Squeeze the bulb to<br />

dispense one hanging drop of<br />

blood.<br />

Do not touch the test strip,<br />

move the meter, or add more<br />

blood during testing.<br />