Error: Test Strip Interference<br />

The test strip was touched or<br />

removed during the test.<br />

Solution:<br />

Power the meter off and<br />

remove the test strip. Repeat<br />

the test using a new test<br />

strip and blood taken from<br />

a new fingerstick from a<br />

different finger.<br />

Do not touch or remove<br />

the test strip when a test is<br />

in progress.<br />

Error: Quality Control Failure<br />

The test strip failed the<br />

internal quality control check.<br />

The test strip is unusable.<br />

Solution:<br />

Power the meter off and<br />

remove the test strip. Repeat<br />

the test using a new test<br />

strip and blood taken from<br />

a new fingerstick from a<br />

different finger.<br />