Error Messages You May See<br />
After You Apply Blood<br />
Error: Time Exceeded<br />
You did not apply blood to<br />
the test strip within 2 minutes<br />
after the blood drop symbol<br />
appeared.<br />
Solution:<br />
Power the meter off and<br />
remove the test strip. Repeat<br />
the test using the same test<br />
strip and blood taken from<br />
a new fingerstick from a<br />
different finger.<br />
Error: Blood Application<br />
Error applying blood to the<br />
test strip.<br />
Solution:<br />
Power the meter off and<br />
remove the test strip.<br />
Repeat the test using a new<br />
test strip and blood taken<br />
from a new fingerstick from a<br />
different finger.<br />
44