View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
View Actual EPA Method 332 (PDF File) - Columbia Analytical ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
[(Dionex AG16, 2-mm internal diameter (I.D.), Part No. 55379) or (Metrohm<br />
ASUPP-4/5 guard, 4-mm I.D., Metrohm Part No. 6.1006.500) or equivalent].<br />
6.1.6 ANION S<strong>EPA</strong>RATOR COLUMN - A 100-250 mm column packed with a solid<br />
phase specially engineered to achieve separation of the anions of interest<br />
[(Dionex AS16, 2-mm I.D. X 250-mm length, Part No. 55378) or (Metrohm<br />
ASUPP5-100, 4-mm I.D. X 100-mm length, Part No. 6.1006.510) or<br />
equivalent].<br />
6.1.7 CONDUCTIVITY SUPPRESSOR - An electrolytic suppressor operated with an<br />
external source of RW. A chemical conductivity suppressor is acceptable,<br />
although sulfuric acid should not be used as the chemical regenerant due to mass<br />
spectrometric interferences caused by HSO 4<br />
-<br />
at m/z 99 [(Dionex Anion Self<br />
Regenerating Suppressor ASRS-MS, 2-mm, Part No. 63008) or (Metrohm<br />
Advanced IC Separation Center, Metrohm Model No. 820, Part No. 2.820.0220<br />
and 2.833.0010) or equivalent].<br />
6.1.8 CONDUCTIVITY DETECTOR - A flow-through detector with an internal<br />
volume that does not introduce analyte band broadening [(Dionex Conductivity<br />
Detector, Model CD25A) or (Metrohm Advanced IC Conductivity Detector,<br />
Metrohm, Model 819, Part No. 2.819.0010) or equivalent].<br />
6.1.9 SAMPLE LOOPS - 50 to 200 µL size. A 200 µL size was used to generate the<br />
data presented in this method. Smaller or larger injection volumes may be used<br />
as long as the Initial Demonstration of Capability (Sect. 9.2), and all calibrations<br />
and sample analyses are performed using the same injection volume.<br />
6.1.10 DATA SYSTEM - Data management software differs from vendor to vendor<br />
and may be recommended by the supplier of the IC or MS. A system that allows<br />
control of both the IC and MS is recommended [(Dionex Chromeleon<br />
Chromatography Management Software, Version 6.4 MSQ) or (Metrohm ICNet<br />
2.3 data management software and Agilent LCMS Chemstation, Metrohm,<br />
v10.02, Part No. G2710AA) or equivalent].<br />
6.1.11 HELIUM - High purity, compressed gas with a pressure of at least 80 psi to<br />
activate valves, sparge eluent and deliver water to the suppressor.<br />
6.1.12 MASS SPECTROMETER - MS equipped with an ESI interface. Operated in<br />
SIM mode [(Dionex Model MSQ-ELMO, manufactured by Thermo Electron,<br />
San Jose, CA) or (Agilent 1100 Series MSD Quad SL, Part No. G1956B,<br />
manufactured by Agilent Technologies, Wilmington, DE) or equivalent].<br />
6.1.13 NITROGEN - Compressed gas for ESI operation, 80 psi. The purity should be<br />
consistent with the MS manufacturer’s recommendations. Due to the high flow<br />
<strong>332</strong>.0-10