equivalent). The latter can be used directly with a sterile vacuum filter unit if not using<br />

syringe filtration.<br />

6.5 SAMPLE FILTERS - Sterile, single-use, disposable surfactant-free cellulose acetate<br />

(SFCA) 26 mm, 0.2 µm syringe filter (Fisher Scientific, Suwanee, GA, Corning Brand,<br />

Part No. 09-754-13 or equivalent). For samples high in particulates, filters with built-in<br />

prefilters are available. All samples must be filtered at the time of sample collection.<br />

6.6 SYRINGES - Sterile, single-use, disposable, silicone-free, luer-lok, 20 mL (Fisher<br />

Scientific, Suwanee, GA, Target Brand, Part No. 03-377-30 or equivalent).<br />

6.7 SAMPLE PRETREATMENT CARTRIDGES - Single-use, disposable OnGuard-II H<br />

cartridges (Dionex, Part No. 057085 or equivalent) used to remove high concentrations<br />

of carbonate if it is determined to be an interferant. OnGuard-II Ba 2+ cartridges<br />

(Dionex, Part No. 57093 or equivalent) used to remove high concentrations of sulfate if<br />

it is determined to be an interferant. OnGuard-II Ag cartridges (Dionex, Part No. 57089<br />

or equivalent) used to remove high concentrations of chloride if it is determined to be an<br />

interferant. The Ba 2+ pretreatment cartridge is the only one that may be required to meet<br />

the QC requirements of this method (Sect.11.6.2)<br />

6.8 MICRO-PIPETTES - 250 µL, 1000 µL and 10 mL sizes with single-use disposable tips<br />

(Rainin, Oakland, CA, Part Nos. EP-250, EP-1000, and EP-10 mL or equivalent).<br />

6.9 VIALS - Single use, disposable autosampler vials with filter caps, or other disposable,<br />

single use vials with caps having a 10 mL or less capacity to be used for sample<br />

preparation.<br />

7. REAGENTS AND STANDARDS<br />

7.1 REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used.<br />

Unless otherwise indicated, it is intended that all reagents shall conform to the<br />

specifications of the Committee on <strong>Analytical</strong> Reagents of the American Chemical<br />

Society (ACS), where such specifications are available. Solvents should be HPLC<br />

grade or better. Other grades may be used provided it is first determined that the reagent<br />

is of sufficiently high purity to permit its use without lessening the quality of the<br />

determination.<br />

7.1.1 HIGH PURITY REAGENT WATER (RW) - Purified water which does not<br />

contain any measurable quantity of the target analyte or interfering compounds<br />

at concentrations > 1/3 the MRL for the target analyte. The purity of the water<br />

required for this method cannot be overly emphasized. For this work, deionized<br />

water was further purified using a bench model Millipore water purification<br />

system (Millipore Corp, Billerica, MA, Model No. MilliQ Gradient A10 or<br />

equivalent).<br />

<strong>332</strong>.0-12