or other interferences in the LRB must be < 1/3 the MRL. If this criterion is not<br />
met, then all data must be considered invalid for all samples in the Analysis<br />
Batch.<br />
NOTE: If samples are collected using devices that have not been previously<br />
evaluated by the laboratory, duplicates of the sample collection devices must be<br />
sent with the samples so an LRB (and an LFB) may be processed in the<br />
laboratory.<br />
NOTE: Although quantitative data below the MRL may not be reliably accurate<br />
enough for data reporting, such data is useful in determining the magnitude of a<br />
background interference. Therefore, blank contamination levels may be<br />
estimated by extrapolation when the concentration is below the lowest<br />
calibration standard<br />
9.3.2 CONTINUING CALIBRATION CHECK (CCC) - CCCs are analyzed at the<br />
beginning of each Analysis Batch, after every ten field samples, and at the end of<br />
the Analysis Batch. See Section 10.4 for concentration requirements and<br />
acceptance criteria.<br />
9.3.3 LABORATORY FORTIFIED BLANK (LFB) - An LFB is required with each<br />
Analysis Batch. The fortified concentration of the LFB must be rotated between<br />
low, medium, and high concentrations from batch to batch. The low<br />
concentration LFB must be as near as practical to the MRL. Similarly, the high<br />
concentration LFB should be near the high end of the calibration range<br />
established during the Initial Calibration (Sect. 10.3). Results of LFBs fortified<br />
at concentrations < the MRL must be recovered within 50-150% of the true<br />
value. Results from the analysis at any other concentration must be recovered<br />
within 80-120% of the true value. If the LFB results do not meet these criteria,<br />
then all data must be considered invalid for all field samples in the Analysis<br />
Batch.<br />
NOTE: LFBs must be processed in the same manner as field samples including<br />
all sample preservation and pretreatment requirements (i.e., sterile filtration) as<br />
described in Section 8.1.4.1.<br />
LFB Fortified Concentration Range<br />
LFB Recovery Limits<br />
< MRL 50 - 150%<br />
>MRL to highest calibration standard 80 - 120%<br />
9.3.4 INTERNAL STANDARD (IS) – The analyst must monitor the peak area of the<br />
internal standard in all injections during each Analysis Batch. The IS response<br />
(as indicated by peak area) for any chromatographic run must not deviate by<br />
<strong>332</strong>.0-20