Application procedure I<br />

Valve functions<br />

Attaching and changing the sole<br />

(1) (2)<br />

Bean bag becomes s<strong>of</strong>t:<br />

Pull valve rim, then bend<br />

valve gently sideways<br />

guiding line<br />

2.)<br />

1.)<br />

Open valve on vacuum cushion (see also valve functions)<br />

and remove the front shell. Place foot in inner boot and<br />

close Velcro tabs. Then insert foot with inner boot in shell.<br />

Application procedure II<br />

Bean bag fixed in place:<br />

Push down the valve rim<br />

and <strong>use</strong> the pump<br />

Removal: Pull down both levers at the same time to<br />

unlock (1) and the push the sole toward the back to<br />

remove (2).<br />

Insert front (ventral) adapter<br />

• Insert adapter on one side (top and bottom at<br />

the same time, see diagram)<br />

• Pull rubber and insert in the same way<br />

on the other side<br />

• For freedom <strong>of</strong> movement at ROM 10° - 0° - 10°,<br />

remove front adapter and do not replace<br />

Insert spreader (1)<br />

In cases <strong>of</strong> a severely swollen ankle, the spreader creates<br />

more space in the rear (dorsal) shell.<br />

• Attach spreader (as illustrated) to the guiding line<br />

on the dorsal shell<br />

• Press the center <strong>of</strong> the spreader inwards with both<br />

hands and assure that the spreader clicks into place<br />

Close valve cap<br />

2.)<br />

Replacement: Do not depress side levers when replacing<br />

sole. Align the raised plastic barbs on the foot sole to the<br />

open slots <strong>of</strong> the boot. Push the sole forward and an<br />

audible click will be heard. Inspect that all the connectors<br />

are secured.<br />

1.)<br />

Use <strong>of</strong> Achilles pad (VACO ® achill only)<br />

Remove spreader (2)<br />

• Press the locking spring inwards on one side with the<br />

thumb, keeping the index and middle fingers on the<br />

outer edge <strong>of</strong> the spreader<br />

• Repeat this procedure on the other side<br />

• Remove spreader<br />

Note: In the US, the ankle spreader is sold separately.<br />

Put on front shell and close straps. The recommended<br />

sequence is as shown in the diagram. For optimum adjustment,<br />

patients who are allowed weight bearing should take<br />

a few steps and then re-adjust the tension <strong>of</strong> the straps.<br />

Application procedure III<br />

Attach the vacuum pump to the valve stem and<br />

evacuate the air. The vacuum cushion will feel hard to the<br />

touch. Remove pump and close valve cap. Check and if<br />

necessary re-adjust the tension <strong>of</strong> the straps.<br />

Exchange <strong>of</strong> terry cloth liner<br />

Opening<br />

for valve<br />

The vacuum cushion is placed inside a terry cloth liner<br />

similar to a pillow inside a case. To get it out, open the<br />

zipper in the back and pull out the vacuum cushion. When<br />

getting the vacuum cushion back into a liner, please<br />

assure that all the wings are inserted properly. Evacuating<br />

the vacuum cushion will make insertion easier. Close the<br />

zipper all the way.<br />

Adapter<br />

The rear adapter is <strong>use</strong>d to change the position <strong>of</strong> the<br />

boot to 0°, 15° or 30°. When the adapter is placed in the<br />

boot the number 0° is at the top and the 30° at the<br />

bottom. The adapter screw has two openings for the key.<br />

Turn the key, positioning the arrow to the open position<br />

and remove the adapter. Adjust the boot by moving the<br />

back (calf) portion. Once in the correct position, insert the<br />

key, turn until the arrow on the adapter points to the<br />

lock position. When the rear adapter is set in a neutral or<br />

0° position, the 0° band on the front shell should be <strong>use</strong>d.<br />

It is recommended to insert the Achilles Pad while using<br />

the VACO ® achill in the 30° plantar flexion position. Place<br />

the Achilles Pad on the inside <strong>of</strong> the posterior rigid shell.<br />

Slide the top portion <strong>of</strong> Achilles Pad under the boot strap,<br />

then snap the hooks in place.<br />

Important safety <strong>instructions</strong><br />

If any pain, numbness in the toes or blueness <strong>of</strong> the<br />

feet cannot be corrected by repeating the adjustment<br />

procedure, see a doctor immediately. Medical<br />

treatment is required in these circumstances.<br />

Always check that the sole is firmly in place and<br />

never walk without a sole – DANGER OF SLIPPING! Do<br />

not walk without front shell being properly attached<br />

and secured by girth straps.<br />

• It is not recommended that patients drive an<br />

automobile while wearing the VACO ® ped<br />

• Ensure that all girth straps are securely tightened<br />

• Inspect the device prior to application for tears,<br />

holes or other damage, particularly after washing.<br />

If you suspect that a device is damaged, contact<br />

your OPED representative