9<br />

DIMENSIONS<br />

Model EJX510A and EJX530A<br />

With process connections code 7<br />

Unit: mm (Approx. inch)<br />

95(3.74)<br />

60(2.36) *2<br />

188(7.4)<br />

External indicator<br />

Conduit connection<br />

(optional)<br />

54<br />

(2.13)<br />

Conduit<br />

connection<br />

Zero<br />

adjustment<br />

110(4.33)<br />

12<br />

(0.47)<br />

39<br />

(1.54)<br />

6<br />

(0.23)<br />

Electrical connection<br />

for code 5 and 9.<br />

129(5.08)<br />

Integral<br />

indicator<br />

(optional)<br />

163(6.42)<br />

Shrouding bolt<br />

(for flame-proof type)<br />

169(6.65)<br />

ø77(3.03)<br />

Ground<br />

terminal<br />

ø69<br />

(2.72)<br />

110(4.33)<br />

<strong>Mount</strong>ing<br />

bracket<br />

(optional)<br />

12 *3<br />

(0.47)<br />

2-inch pipe<br />

(O.D. 60.5mm)<br />

Pipe for<br />

open to<br />

atmosphere *1<br />

129(5.08)<br />

*1: Only for EJX530A whose measurement span code is A, B, or C.<br />

*2: 58 mm (2.28 inch) for measurement span code D.<br />

*3: 11 mm (0.43 inch) for measurement span code D.<br />

With process connections code 4<br />

With process connections code 8 and 9<br />

20(0.78)<br />

174(6.85) 158(6.22)<br />

5(0.20)<br />

ø6(0.23)<br />

F04.EPS<br />

● Terminal Configuration<br />

Communication<br />

terminals (BT200 etc.)<br />

connection hook<br />

Check meter<br />

connection hook *4*5<br />

● Terminal Wiring<br />

SUPPLY<br />

CHECK<br />

or<br />

ALARM<br />

+<br />

–<br />

+<br />

–<br />

+<br />

–<br />

Power supply and output terminal<br />

External indicator (ammeter) terminal *4*5<br />

or<br />

Status contact output terminal *5<br />

(when /AL is specified)<br />

SUPPLY +<br />

CHECK + or<br />

ALARM +<br />

SUPPLY – CHECK – or ALARM –<br />

Ground terminal<br />

*4: When using an external indicator or check meter, the internal<br />

resistance must be 10 or less. A check meter or indicator<br />

cannot be connected when /AL option is specified.<br />

*5: Not available for fieldbus communication type.<br />

F05.EPS<br />

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation<br />

GS 01C25F01-01E Feb. 10, 2006-00