18.11.2012 Views

The 2D-ARRAY seven29 A new way of dosimetric verification of - PTW

The 2D-ARRAY seven29 A new way of dosimetric verification of - PTW

The 2D-ARRAY seven29 A new way of dosimetric verification of - PTW

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

• By default, VeriS<strong>of</strong>t normalizes each matrix individually to its maximum dose. This is not<br />

meaningful for absolute dose measurements with the <strong>2D</strong>-<strong>ARRAY</strong>. <strong>The</strong>refore, both data sets<br />

have to be normalized to the same meaningful value (e.g., the dose at the isocenter).<br />

• <strong>The</strong> correct positioning <strong>of</strong> the two data sets is checked using the “Pr<strong>of</strong>iles” option. Normally, no<br />

alignment should be required. However, positional corrections <strong>of</strong> ±1 mm in both x and y<br />

directions are sometimes necessary due to slight misalignments <strong>of</strong> the phantom during the<br />

measurements.<br />

Fig. 13a shows the isodose overlay, which already gives a first impression <strong>of</strong> the concurrence <strong>of</strong><br />

the two data sets. This concurrence is quantitatively confirmed by vertical and horizontal line scans<br />

through the isocenter (Fig. 13b).<br />

Line scans have the disadvantage that an investigation <strong>of</strong> the complete area is quite timeconsuming.<br />

<strong>The</strong>refore, VeriS<strong>of</strong>t provides additional quantitative <strong>2D</strong> compare modes.<br />

Figure 13a: VeriS<strong>of</strong>t isodose overlay <strong>of</strong> a typical IMRT<br />

head and neck field. <strong>The</strong> measured isodoses are depicted<br />

as dotted lines.<br />

In Fig. 14, a difference plot is shown, where each measured chamber dose Dm(i,j) is compared to<br />

the averaged calculated dose according to the following formula:<br />

∆D(<br />

i,<br />

j)<br />

D<br />

D<br />

=<br />

m<br />

( i,<br />

j)<br />

D<br />

−<br />

c<br />

D<br />

c<br />

( i,<br />

j)<br />

( i,<br />

j)<br />

* 100%<br />

Note that VeriS<strong>of</strong>t provides also a percentage difference mode, which is relative to the<br />

normalization dose <strong>of</strong> matrix A. This mode however is not as sensitive to the low dose.<br />

10 / 16<br />

Figure 13b: Horizontal (above) and vertical (below)<br />

line scans through the isocenter <strong>of</strong> the IMRT field <strong>of</strong> Fig. 13a.<br />

Calculations are shown as blue lines and the measured<br />

chamber values are depicted as green bars.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!