NS5<strong>00</strong>0/6<strong>00</strong>0 <strong>Handler</strong> Interface Manual for the Catalyst Test SystemModifying the <strong>Handler</strong> Base Plate (If Necessary)Complete this section if it is necessary to add four 1/4"-20 UNC-2 holes to the handler baseplate for mounting the interface plate. See Figure 3.If the holes already exist, go to: Installation on page 13.1) Loosen the four fasteners that hold the two fixture plates to the large base plate (seeFigure 3 and Figure 4).2) Orient the drill jig with the alignment pins facing up and oriented as shown in Figure 3.Note the location of large hole on the drill jig..The drill jig (pn 480-073-<strong>00</strong>) uses some of the holes in the large base plate. Align thealignment pins of the drill jig to the holes shown in Figure 5.Fixture Plate (X2)Small Base PlateLarge Base PlateLarge HoleDrill Jig, NS 5K/6K(PN 480-073-<strong>00</strong>)Small HoleAlignment Pins (X4)Figure 3 Drill Jig Alignment9