P1.1 iCAL – Internal calibrationstart of internal calibration process, process is completely automatic withoutinterference of balance operator, if pan is loaded display will indicate messageabout necessity of removing the load.or CAL buttonP1.2 ECAL – External calibrationcalibration with external weight, which value is saved in factory menu ofbalance, function unavailable in verified balances.balances with internal calibrationbalances with external calibrationP1.3 tCAL - Test of calibrationcomparison of internal calibration mass with its value saved in balance memory.balances with internal calibrationbalances with external calibrationP1.4 ACAL – Automatic calibration (balances with internal calibration)determination of factor, which should decide about starting the automatic internalcalibrationnonE – none of factors will cause start of calibrationtEmP – calibration with regard to change of temperaturetimE – calibration with regard to time set in P1.5 CALtboth – calibration with regard to time and temperature.34