<strong>SensoCard</strong>10. E-9 error message• External temperature is too high or low (out <strong>of</strong> the range 10-40 °C). Wait at least 30minutes before using the device, until it completely reaches room temperature.<strong>SensoCard</strong> is able to check external temperature. If it detects values not conforming toits operational requirements, the following error messages will be displayed:• An arrow icon directed downward and the E-9 error message together indicate lowtemperature (below 10 °C).• An arrow icon directed upward and the E-9 error message together indicate hightemperature (above 40 °C).11. E-C error messageThis error message may appear during using Code-card. There might be one <strong>of</strong> the followingproblems:• Code-card is damaged. Please check the integrity <strong>of</strong> Code-card. If you can find anymechanical damage on Code-card, contact your local distributor.• Code-card insertion or removal was not continuous. Insert and remove Code-cardwith a bit quicker expressed move.52