4.8.3 Clear ZeroThis will clear the zero offset and return to the calibrated absolute output.4.8.4 3 Pt. AdjustThe 3 point adjust makes a slight adjustment to an existing calibration. It allows for correction ofinstallation or slight loading errors in gain and offset. For this calibration, the sensor must bepositioned at MIN, MID, and MAX with data taken at each point. This adjustment does notoptimize the output further; it simply corrects scale, offset, and small linearity errors.Temperature coefficients are not changed. It will remove any offset from zeroing the sensoroutput. This option is used when more accuracy, but not a full calibration is required.www.kaman<strong>sensors</strong>.com PART NO: 860525-001Last Revised 9/16/2013- 12 -