MAIN MENUAUTO SAMPLEMANUAL SAMPLECALIBRATION>>>CALIBRATIONSPAN CALIBRATEZERO CALIBRATEDEFAULT SPANDEFAULT ZEROOUTPUT SPANOUTPUT ZERO17. Press the ENTER key to select the SPAN CALIBRATE option.18. Note: A span gas concentration above 1000 <strong>ppm</strong> dictates the selection of the PERCENT option.19. Advance the reverse shade cursor using the ARROW keys to highlight the desired GAS CONCENTRATION.20. Press the ENTER key to select the highlighted menu option.GAS CONCENTRATIONPERCENTPPM21. The following displays appear:000.00 PPMPRESS UP OR DOWNTO CHANGE VALUEENTER TO SAVEMENU TO RETURN>>>080.00 PPMSPANCALIBRATIONENTER TO CALIBRATEMENU TO ABORT22. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left toreach to the desired digit of the alarm value.23. Press the ARROW keys to enter the alarm value.24. Repeat until the complete SPAN value has been entered.25. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.26. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output orreading has stabilized within 60% of the full scale low range. Both the Zero Calibrate and Span Calibrate functions result inthe following displays:PASSEDCALIBRATIONORFAILEDCALIBRATION27. If the calibration is successful, the analyzer returns to the SAMPLING mode after 30 seconds.28. If the calibration is unsuccessful, return to the SAMPLING mode with span gas flowing through the analyzer, make sure thereading stabilizes and repeat the calibration before concluding the equipment is defective.29. Before disconnecting the span gas line, place the SAMPLE/BYPASS valve in the BYPASS position.22

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!