Microprocessor Digital <strong>Manifold</strong>SYMPTOM CAUSE CORRECTIVE ACTIONShuttle Valve FaultsGas leakage.a. Body halves not joinedtightly enough.1. Tighten screws.b. Body “O” ring leaking.c. Gas bleeding out of shaftof shuttle valve, (i.e.: leftside and feeding from theright high pressure side).d. Shaft “O” ring leaking, (i.e.:left side, and feeding fromthe left high pressure side).2. Replace “O” ring kit, userepair kit.3. Replace check-valve(s).4. Use shuttle valve repair kit.Both banks feeding.a. Shuttle valve seat leaking.b. Shuttle valve shaft hungup.1. Replace shuttle valve seal,use repair kit.2. Bleed gas, reset operatingregulators.ERROR CODE MESSAGES ON THE MANIFOLD DISPLAYSYMPTOM CAUSE CORRECTIVE ACTIONE01Sensor not connected.Check sensor wiring. Reconnectwire properly.E06Sensor wires between themanifold module, shorted outor reversed polarity.Reverse polarity or replacedefective wire.E10Shuttle valve shaft hung up ornot shuttled-fully. Operatingregulators not set up properly,limit switch wiring connector.Repair shuttle valve by usingrepair kit.Reset operating regulators tospecified setting, make sureconnections are properly connectedto switches.Note: Under Low Flow conditions, E10 could happen during switchover. This is normal under LowFlow condition. When the shuttle has fully shuttled over the E10 error will clear.Page: 26

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!