Microprocessor Digital <strong>Manifold</strong>APPENDIX - FWiring Schematic for Heater UnitsHeaterCartridge WiresHeaterTerminal BlockHeaterCartridge WiresReset-SwitchInput PressureInput PressurePlease note that power must be connected to the terminal blockshown in the above diagram in order for the heaters to function.The heaters will turn on when the temperature drops below 24 deg.C. If the temperature exceeds 65-75 deg. C. the heater re-setswitch will trip and the heaters will automatically turn off. Heatersare used on Nitrous Oxide & Carbon Dioxide <strong>Manifold</strong>s only. To resetthe heaters, remove the heater covers, press the red button onthe re-set switch to activate.Page: 36