54|55<br />

Likewise, a decrease in RVR can indicate a success of weaning. Then we<br />

probably have to decrease the pressure support level in order to shift a further<br />

part of the WOB from the ventilator to the patient.<br />

RVR (↓)<br />

too less WOB <strong>for</strong> the patient<br />

= training success<br />

Pinsp (↓)<br />

The Rate Volume Ratio is monitored by the Babylog 8000plus and can be<br />

displayed as value and graphical trend. Starting from the main menu, press<br />

>Trend< and >Param< repeatedly until the parameter RVR appears on<br />

screen.<br />

D-22252-2010<br />

Up to now there are no studies available, which focus on this new parameter<br />

<strong>for</strong> use in neonatal <strong>Pressure</strong> <strong>Support</strong> <strong>Ventilation</strong>. Nevertheless the Rate<br />