11.12.2012 Views

Anritsu Site Master S332B User Guide - Mr Test Equipment

Anritsu Site Master S332B User Guide - Mr Test Equipment

Anritsu Site Master S332B User Guide - Mr Test Equipment

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Symbols<br />

Table 3-1 provides a listing of the symbols used as condition indicators on the LCD display.<br />

Table 3-1. LCD Icon Symbols<br />

Icon Symbol<br />

HOLD<br />

�dx<br />

T<br />

�<br />

Self <strong>Test</strong><br />

Chapter 3 Functions and Operations<br />

<strong>Site</strong> <strong>Master</strong> is in Hold for power conservation. To resume sweeping, press<br />

the RUN/HOLD key. After 10 minutes without a key press, the <strong>Site</strong> <strong>Master</strong><br />

will automatically activate the power conservation mode.<br />

Integrator Failure<br />

Lock fail indication. Check battery. (If the <strong>Site</strong> <strong>Master</strong> fails to lock with a<br />

fully charged battery, call your <strong>Anritsu</strong> Service Center.)<br />

When calibration is performed, the <strong>Site</strong> <strong>Master</strong> stores the ambient temperature.<br />

If the temperature drifts outside the specified range, this indicator<br />

will flash. A recalibration at the current temperature is recommended.<br />

Indicates the remaining charge on the battery. The inner white rectangle<br />

grows longer as the battery charge depletes.<br />

Indicates internal data processing.<br />

At turn-on, the <strong>Site</strong> <strong>Master</strong> runs through a series of quick checks to ensure the system is<br />

functioning properly. Note that the battery voltage and temperature are displayed in the<br />

lower left corner below the self test message. If the battery is low, or if the ambient temperature<br />

is not within the specified operational range, Self <strong>Test</strong> will fail. If Self <strong>Test</strong> fails and<br />

the battery is fully charged and the <strong>Site</strong> <strong>Master</strong> is within the specified operating range, call<br />

your <strong>Anritsu</strong> Service Center.<br />

3-21

Hooray! Your file is uploaded and ready to be published.