24 Infant Flow ® SiPAP<br />

Electronics Module<br />

The function of the electronics module is to power the unit either by AC mains supply<br />

or DC emergency battery supply, to control the gas module and read the gas module<br />

sensors. The main components are a power supply unit, a rechargeable battery, a<br />

main processor PCB, LED PCB, Valve/Sensor PCB and a LCD screen (touch<br />

screen). The LCD screen includes a back-light which is always on when the IFSD is<br />

powered. The touch screen displays information and receives inputs from the Operator<br />

via the touch screen keyboard.<br />

Figure 2 - IFSD Back Panel<br />

Ground test<br />

point<br />

675-120 Revision C November 2004