18.01.2013 Views

Service Manual - Humatem

Service Manual - Humatem

Service Manual - Humatem

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Service</strong> <strong>Manual</strong> 36<br />

Chapter 6 - Operation<br />

The operating procedures below show the procedures for all models. Reference<br />

should only be made to the procedures for the model in use. Read the Warnings and<br />

Cautions at the beginning of this manual before you start the procedures.<br />

Preparing and Connecting the Equipment<br />

WARNING<br />

Ensure the whistle sounds during gas connection, if not the device may be<br />

faulty and should not be used.<br />

1 Connect the Medical Air and Oxygen hoses to the IFSD connections on the<br />

back panel.<br />

Note<br />

The gas failure alarm/bypass warning whistle sounds until the pressures balance.<br />

WARNING<br />

Only use the supplied AC cable to connect to the power supply.<br />

2 Connect the power supply cable to a suitable power supply outlet. The green<br />

power light will come on regardless of the position of the power switch.<br />

3 Connect the patient circuit to the IFSD as required. Figure 9 shows a typical<br />

configuration, the actual configuration may vary dependent on the type of ancillary<br />

equipment used and the clinical needs prescribed for the patient. For functional<br />

test purposes, the Generator nasal prongs can be occluded to simulate patient<br />

responses.<br />

4 Connect the transducer module to the sensor and to the IFSD. For functional<br />

test purposes, lightly tap the abdominal respiratory sensor to simulate patient<br />

breathing.<br />

675-120 Revision C November 2004

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!