50 Infant Flow ® SiPAP<br />

Alarm Method of Setting<br />

Software not running with<br />

unit connected to power<br />

Blender whistle Water trap blocked,<br />

full or leaking; filters<br />

blocked; loss of wall<br />

pressure; imbalance<br />

of wall gas supply.<br />

Oxygen cell calibration<br />

error.<br />

Indications and<br />

Actions<br />

Automatic Constant Audible<br />

Alarm.<br />

Warning Bar on.<br />

Screen displays<br />

flashing fault code (E<br />

number).<br />

Automatic monitoring<br />

(oxygen cell<br />

incorrectly calibrated,<br />

damaged or depleted).<br />

Method of Resetting<br />

or Canceling<br />

Cannot be reset.<br />

Refer to <strong>Service</strong><br />

Engineer.<br />

Audible Alarm. Refer to <strong>Service</strong><br />

Engineer.<br />

Difference between<br />

displayed and set<br />

value.<br />

Flowmeter fault. Flowmeter fault No flow indications or<br />

flow cannot be<br />

adjusted.<br />

Electrical fault. Electrical fault External power light<br />

does not match screen<br />

icon.<br />

Calibrate or replace<br />

oxygen cell. Refer to<br />

<strong>Service</strong> <strong>Manual</strong>.<br />

Refer to <strong>Service</strong><br />

Engineer.<br />

Refer to <strong>Service</strong><br />

Engineer.<br />

675-120 Revision C November 2004