38 Infant Flow ® SiPAP<br />

5 Make sure that the water trap is empty. If necessary, empty any water from the<br />

trap (refer to Chapter 3, Maintenance).<br />

6 Gas Flow Pressure Setting<br />

The IFSD provides a virtually constant airway pressure irrespective of patient<br />

demand or expiratory flows via the specially designed generator and nasal prongs. This<br />

is the reason for the IFSDs ability to provided superior NCPAP. The IFSD is subject<br />

to a direct relationship between controlled enriched gas flow and NCPAP pressure.<br />

A nomogram illustrating the relationship between constant airway pressure and flow<br />

settings is shown in Figure 10. Example: 8 L/min gas flow provides 5 cm H2O<br />

NCPAP.<br />

Note<br />

Individual devices have a tolerance of up to ± 10% from that illustrated in the<br />

nomogram and in particular, at pressures below 2 cm H2O.<br />

Figure 10 - Flow Pressure Nomogram<br />

675-120 Revision C November 2004