28 Infant Flow ® SiPAP<br />
• an audible Whistle which sounds a constant tone when the gas imbalance<br />
limits are exceeded.<br />
• an audible Alarm which sounds a constant tone or two different intermittent<br />
tones; medium or high level (medium level sounds beep.beep.beep every 15<br />
seconds and high level sounds beep.beep..beep beep beep every 10 seconds).<br />
• a visual Warning Bar which flashes red continuously, flashes intermittently or<br />
comes on continuously.<br />
• an Alarm Button on the touch screen status bar which flashes continuously.<br />
Other indications are given to indicate the status of the condition. Refer to Table 2 for<br />
alarms and indications for specified conditions.<br />
Diagnostic Screen<br />
A diagnostic screen is provided for low level calibration, test procedures for valves<br />
and an error log. Figure 5 shows the screen display. Labeling of Diagnostic screen<br />
will be the same regardless of configuration.<br />
Figure 5 - Diagnostic Screen<br />
675-120 Revision C November 2004<br />
-