4.3 Start the creep test.<br />

4.4 Allow the console time to stabilize at the creep pressure. (Small fluctuations in<br />

pressure are due to the lack of volume in the pressure manifold.)<br />

5.0 Verification of Calibration<br />

5.1 After reaching a stable creep pressure verify the calibration of the <strong>2600</strong> by<br />

comparing the pressures displayed on the <strong>2600</strong> console and the calibration<br />

standard (Figures 5 and 6).<br />

5.2 The pressures should be within 2.77 in H2O (0.1 psi) of each other.<br />

5.3 Disconnect the Calibration Standard, associated tubing and the Pressure Manifold.<br />