Calibration Check<br />

The <strong>Micro</strong> I is calibrated to read in litres at body temperature,<br />

barometric pressure saturated with water vapour (BTPS).<br />

The calibration should remain stable indefinitely, unless the<br />

transducer is physically damaged, and the unit should not require<br />

re-calibration. However, to ensure the correct functioning of the<br />

unit we do recommend that a calibration check is performed<br />

periodically.<br />

To perform a calibration check select calibration check option from<br />

the setting menu and the display will show:<br />

Connect a 3L syringe to the <strong>Micro</strong> I with the minimum of adapters<br />

and inject the syringe volume into the transducer evenly, without<br />

pausing. When complete the display will show:<br />

The acceptable calibration accuracy is +/- 3.5%. If the calibration<br />

accuracy is outside of this range then the tick will be replaced by a<br />

cross. If this happens then check the syringe and the connections<br />

for leaks and repeat the check. If the <strong>Micro</strong> I is repeatedly outside<br />

30