To aid this process clip 1 has a recess moulded into the device body<br />

to allow a small flat bladed screw driver to be carefully slid under<br />

the clip to release it.<br />

Once released work the screw driver around to clips 2, 3 and 4.<br />

Finally slide the cover off the bottom of the unit which in turn<br />

releases clips 5 and 6<br />

Replace the battery paying <strong>part</strong>icular attention to the orientation of<br />

the connector in its socket and of the battery itself in the case. Note<br />

the leads coming from the battery should exit from the upper face of<br />

the pack when installed.<br />

Replace the rear cover by attaching the securing clips in the reverse<br />

order.<br />

Dispose of the waste battery in line with EU Battery Regulations.<br />