3. DESCRIPTION<br />
See 15 - Attachment<br />
Functions common to <strong>F11N</strong> and <strong>F13N</strong><br />
Clamping the cable.<br />
The specially designed shape of the jaws enables large crosssection<br />
conductors to be enclosed, and satisfies the majority<br />
of high current applications.<br />
A 2 busbars 50 x 5 mm<br />
B 2 cables Æ 25 mm<br />
C 1 cable Æ 42 mm<br />
Reminder : The clamp must only enclose a single phase<br />
conductor, which can be physically subdivided into 2 smaller<br />
sections. In principle, the clamp cannot measure current in a<br />
conductor that includes the phase cable and the neutral cable.<br />
1 Jaws<br />
They enclose the cable to measure AC CURRENTS.<br />
The jaws are designed with an "anti-pinch" system.<br />
The red colour was chosen to identify the zone which is<br />
potentially dangerous during measurement of AC currents.<br />
2 Guard<br />
The non-slip guard protects the user from accidental contact<br />
with an uninsulated conductor when clamping a cable.<br />
3 Trigger<br />
To open the jaws.<br />
4 MAX button<br />
To activate the max value function on A and on V and for the<br />
diode test function.<br />
Typical acquisition time: < 500 ms (except for calibre change).<br />
NB : This button is inoperative on frequency and resistance<br />
measurement.<br />
5 HOLD buttons<br />
Cette fonction permet de mesurer dans un environnement<br />
difficile, ou peu éclairé, en mémorisant la valeur pour une<br />
lecture différée.<br />
Cette fonction est active pour toutes les positions du commutateur<br />
6 .<br />
It is activated by pressing one of the two keys 5 and released<br />
by pressing one of these two keys again (two-way switch<br />
system).<br />