4.6 Resistance<br />
Accidental<br />
W Resolution Accuracy brief overload<br />
400 W 100 mW ± 1 % R ± 5 ct<br />
600 V rms<br />
4000 W 1 W ± 1 % R ± 3 ct<br />
NB : Max open circuit voltage : 4.4 V<br />
Max short circuit current : 1.2 mA<br />
4.7 Continuity sound test<br />
Continuous sound signal for a resistance less than<br />
40 W ± 10 W.<br />
Reponse time: 10 ms.<br />
Max open circuit voltage : 4.4 V<br />
Max short circuit current : 1.2 mA<br />
Permitted accidental overload : 600 V rms.<br />
4.8 Diode test<br />
Measurement of the voltage in volts (V) forward direction,<br />
OL or “voltage > 3.5 V” in reverse direction.<br />
Max open circuit voltage : 4.4 V<br />
Max short circuit current : 1.2 mA<br />
Accuracy : 3 % R + 10 ct, Resolution : 1mV<br />
Permitted accidental overload : 600 V rms.<br />
▼<br />
4.9 Frequency (for the <strong>F13N</strong> only)<br />
Hz<br />
4 kHz<br />
Réesolution<br />
Accuracy<br />
1 Hz ± 0.1 % R ± 1 ct<br />
20<br />
Accidental<br />
brief overload<br />
1000 V rms<br />
or 1000A peak<br />
Trigger threshold: 1 V or 1 A.<br />
Note: frequencies higher than 4 kHz can be indicated, but<br />
only for information purposes.<br />
4.10 Max value (MAX)<br />
Max. value of the positive signal measured.<br />
This function is active for AC currents and DC and AC voltages.<br />
Typical acquisition time : < 500 ms.<br />
Additional error :<br />
■ For the <strong>F11N</strong> :<br />
± 2.5 % R ± 3 ct (high calibre), ± 2.5 % R ± 30 ct (low calibre).<br />
■ For the <strong>F13N</strong> :<br />
Accuracy of the calibre + additional error : ± 2.5 % R ± 5 ct.<br />
4.11 Memorisation (HOLD)<br />
This function records the displayed value and is active for all<br />
the switch positions.