Autoclaving<br />
This instrument is autoclavable at 121 °C (250 °F), 2 bar absolute (30 psi) with a holding time of at least<br />
15 minutes according to DIN EN 285. Before autoclaving, the instrument must be cleaned carefully (page<br />
48-50). Lift the piston out of the cylinder (page 49).For autoclaving with mounted filling tube, the filling valve<br />
with olive-shaped nozzle is recommended (page 55).<br />
1. Prior to the first autoclaving<br />
Remove built-in inelastic PTFE-sealing ring (R) and replace with the<br />
supplied elastic sealing ring made of FKM (S).<br />
Note:<br />
FKM has limited chemical resistance!<br />
2. Autoclaving<br />
a) Loosen screw connections (Fig. a).<br />
b) Pull out the recirculation tube and the telescoping filling tube.<br />
c) Check that the filling valve is securely seated (Fig. c). If the<br />
rigid PFTE ring has been installed, the filling valve must also be<br />
loosened.<br />
In the digital model, check that the piston mount is securely<br />
seated (Fig. c’).<br />
d) Autoclave the instrument with the piston pushed all the way<br />
downward, and with the filling tube and recirculation tube disconnected.<br />
Avoid contact with metallic surfaces. The mounting<br />
tool can also be autoclaved.<br />
Note:<br />
Do not reassemble the instrument until it has cooled down to room<br />
temperature (Cooling time approx. 2 hours).<br />
After every autoclaving, inspect all parts for deformities or damage.<br />
If necessary, replace them.<br />
It is the user's responsibility to ensure effective autoclaving.<br />