Avaliação das alterações geométricas e dimensionais ... - Dentsply
Avaliação das alterações geométricas e dimensionais ... - Dentsply
Avaliação das alterações geométricas e dimensionais ... - Dentsply
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Câmara AS. Evaluation of instruments structural changes from NiTi ProTaper System to NiTi<br />
ProTaper Universal System and their effects over torsional resistance and flexibility [tese<br />
doutorado]. Araraquara: Faculdade de Odontologia da UNESP; 2008.<br />
ABSTRACT<br />
The present study has proposed to evaluate the instruments<br />
structural changes from NiTi ProTaper System to NiTi ProTaper Universal<br />
System as well this effect over torsional resistance and flexibility. It was used a<br />
total amount of 298 instruments. At the beginning, 12 instruments of each size<br />
from both systems were analyzed dimensional and geometrically by optical<br />
microscopy under parameters of angle, length and tip diameter, pitch’s length<br />
along the blade and instrument’s diameter from tip until the end of the active<br />
part. Besides, one sample of each size had its 3-mm from the tip’s cross<br />
sectional area calculated under SEM images. After that, all the instruments<br />
were randomly divided into 4 different groups: Group 1 having 60 ProTaper<br />
instruments inside (S1, S2, F1, F2 and F3),12 samples from each size, to be<br />
submitted to torsion test; Group 2 having 60 ProTaper instruments inside (S1,<br />
S2, F1, F2 and F3),12 samples from each size, to be submitted to bending test;<br />
Group 3 having 84 ProTaper Universal instruments inside (S1, S2, F1, F2, F3,<br />
F4 and F5),12 samples from each size, to be submitted to torsion test; Group 4<br />
having 84 ProTaper instruments inside (S1, S2, F1, F2, F3, F4 and F5),12<br />
samples from each size, to be submitted to bending test. These torsion and<br />
bending tests were performed using a bench test device connected with a<br />
software to acquire and process data specially developed to this purpose,<br />
according to ISO 3630-1. Maximum torque, angular deflection until fracture and<br />
bending moment values were statistically analysed using Student’s t test. The<br />
results have shown that ProTaper and ProTaper Universal system’s instruments<br />
presented good geometric standardization and acceptable surface features,<br />
however differences were detected among the systems. The maximum torque<br />
and the bending moment at 45º. were higher for instruments with larger<br />
diameters and cross sectional area at 3mm from the tip in both systems even<br />
though angular deflection hasn’t shown this same relationship.<br />
Key Words: Endodontics; Torque; Fracture