General Description<br />
Page 10<br />
<strong>LP6</strong> <strong>Plus</strong> and <strong>LP10</strong> User’s Manual<br />
Warning Ensure the proper connection and operation of the patient ventilator<br />
circuit daily. The patient could be at risk if the manifold<br />
does not function as intended. Connecting patient pressure and<br />
exhalation tubes to the incorrect port prevents proper patient<br />
ventilation.<br />
C. Patient Air Hose: This is the large hose between the Bacteria Filter<br />
and the Exhalation Manifold.<br />
D. Patient Pressure Tube (included with patient air hose): This small<br />
tube connects the Patient Pressure port on the ventilator to the Exhalation<br />
Manifold.<br />
E. Exhalation Tube (included with patient air hose): This small tube<br />
connects the Exhalation Valve port to the Exhalation Manifold.<br />
F. Bacteria Filter: This filter cleans the incoming air before you inhale<br />
it.<br />
Other accessories are available. When using any accessory, always follow<br />
the manufacturer’s recommendations and instructions.<br />
Caution Adding attachments or other components to the breathing system will<br />
increase the inspiratory and expiratory resistance.<br />
June 1999