Operating Modes<br />
Page 30<br />
Pressure Limit<br />
(<strong>LP10</strong> Only)<br />
<strong>LP6</strong> <strong>Plus</strong> and <strong>LP10</strong> User’s Manual<br />
The pressure limit valve is a mechanical pressure valve located behind the<br />
Patient Air tube. The ventilator’s microprocessor does not control this<br />
valve.<br />
Note Pressure limit control ventilation is intended for use with uncuffed<br />
tracheostomy tubes, or in other circuit configurations which ensure<br />
an intentional airway leak.<br />
The Pressure Limit control limits the pressure in the airway during inspiration.<br />
Inspiration continues after reaching the preset high pressure level<br />
and lasts until the Inspiratory Time expires.<br />
Use the following instructions to activate and adjust the Pressure Limit<br />
Control.<br />
To set the Pressure Limit level on the <strong>LP10</strong>:<br />
1. Disconnect the patient from the ventilator. Provide another means of<br />
ventilation.<br />
2. Turn the Pressure Limit Control counterclockwise until it stops.<br />
Note The outside ring of the knob must be pushed in before the center adjustment<br />
knob can be moved.<br />
3. Set all controls to the prescribed settings.<br />
4. Block the Exhalation manifold at the port that connects to the patient.<br />
Warning Wash hands thoroughly. Do not introduce germs or contaminants<br />
into the patient circuit while performing this task.<br />
5. Watch the needle on the Patient Pressure Meter. Note the highest<br />
pressure achieved during a machine-delivered breath. This should be a<br />
very low pressure reading.<br />
6. Turn the Pressure Limit Control clockwise in small increments until<br />
the meter needle reaches, but does not exceed, the pressure value prescribed<br />
by the physician.<br />
June 1999