Operating Modes<br />

Page 30<br />

Pressure Limit<br />

(<strong>LP10</strong> Only)<br />

<strong>LP6</strong> <strong>Plus</strong> and <strong>LP10</strong> User’s Manual<br />

The pressure limit valve is a mechanical pressure valve located behind the<br />

Patient Air tube. The ventilator’s microprocessor does not control this<br />

valve.<br />

Note Pressure limit control ventilation is intended for use with uncuffed<br />

tracheostomy tubes, or in other circuit configurations which ensure<br />

an intentional airway leak.<br />

The Pressure Limit control limits the pressure in the airway during inspiration.<br />

Inspiration continues after reaching the preset high pressure level<br />

and lasts until the Inspiratory Time expires.<br />

Use the following instructions to activate and adjust the Pressure Limit<br />

Control.<br />

To set the Pressure Limit level on the <strong>LP10</strong>:<br />

1. Disconnect the patient from the ventilator. Provide another means of<br />

ventilation.<br />

2. Turn the Pressure Limit Control counterclockwise until it stops.<br />

Note The outside ring of the knob must be pushed in before the center adjustment<br />

knob can be moved.<br />

3. Set all controls to the prescribed settings.<br />

4. Block the Exhalation manifold at the port that connects to the patient.<br />

Warning Wash hands thoroughly. Do not introduce germs or contaminants<br />

into the patient circuit while performing this task.<br />

5. Watch the needle on the Patient Pressure Meter. Note the highest<br />

pressure achieved during a machine-delivered breath. This should be a<br />

very low pressure reading.<br />

6. Turn the Pressure Limit Control clockwise in small increments until<br />

the meter needle reaches, but does not exceed, the pressure value prescribed<br />

by the physician.<br />

June 1999