STERILIZATION PROCEDURES<br />
26<br />
Note<br />
Do not place carbon steel instruments<br />
directly onto stainless steel<br />
trays. Place an unbleached towel<br />
or paper wrap on the tray and<br />
place the instruments on the towel<br />
or paper.<br />
When wrapping instruments for<br />
sterilization, use a material which<br />
promotes drying.<br />
Warning<br />
Do not overload the trays.<br />
Overloading trays may result in<br />
inadequate sterilization and<br />
unsterile instruments.<br />
Warning<br />
Monitor the sterilization process with<br />
process monitors and biological indicators<br />
as described in<br />
Recommended Monitoring (p. 33).<br />
Preparation of Carbon Steel<br />
Instruments<br />
Carbon steel instruments must be carefully prepared prior to<br />
sterilization to prevent rusting. After cleaning according to the<br />
manufacturer's recommendations or the methods presented<br />
in Instrument Cleaning Prior To Sterilization, prepare carbon<br />
steel instruments as follows:<br />
1. Prepare a 2% solution of sodium nitrate in water<br />
(approximately one table spoon in a quart of<br />
water). Completely immerse the instruments in the<br />
solution and allow them to soak for three minutes.<br />
2. Remove the instruments and package them for<br />
sterilization or place on trays if sterilizing<br />
unwrapped. Do not rinse or wipe instruments<br />
prior to sterilization.<br />
Preparing Items for Sterilization<br />
General Preparation Guidelines<br />
1. Clean items thoroughly before placing them in the<br />
sterilizer. Follow good laboratory or medical procedures<br />
for cleaning and decontamination of items.<br />
2. Place absorbent paper or linen towel in the bottom<br />
of trays before use.<br />
3. Unwrapped goods should only be processed on<br />
trays. Do not place items on the rack or the bottom<br />
of the sterilizer.<br />
4. Liquids must not exceed 75% of capacity of flask.<br />
Use self-venting caps.<br />
5. Open all hinged instruments.