STERILIZATION PROCEDURES<br />

26<br />

Note<br />

Do not place carbon steel instruments<br />

directly onto stainless steel<br />

trays. Place an unbleached towel<br />

or paper wrap on the tray and<br />

place the instruments on the towel<br />

or paper.<br />

When wrapping instruments for<br />

sterilization, use a material which<br />

promotes drying.<br />

Warning<br />

Do not overload the trays.<br />

Overloading trays may result in<br />

inadequate sterilization and<br />

unsterile instruments.<br />

Warning<br />

Monitor the sterilization process with<br />

process monitors and biological indicators<br />

as described in<br />

Recommended Monitoring (p. 33).<br />

Preparation of Carbon Steel<br />

Instruments<br />

Carbon steel instruments must be carefully prepared prior to<br />

sterilization to prevent rusting. After cleaning according to the<br />

manufacturer's recommendations or the methods presented<br />

in Instrument Cleaning Prior To Sterilization, prepare carbon<br />

steel instruments as follows:<br />

1. Prepare a 2% solution of sodium nitrate in water<br />

(approximately one table spoon in a quart of<br />

water). Completely immerse the instruments in the<br />

solution and allow them to soak for three minutes.<br />

2. Remove the instruments and package them for<br />

sterilization or place on trays if sterilizing<br />

unwrapped. Do not rinse or wipe instruments<br />

prior to sterilization.<br />

Preparing Items for Sterilization<br />

General Preparation Guidelines<br />

1. Clean items thoroughly before placing them in the<br />

sterilizer. Follow good laboratory or medical procedures<br />

for cleaning and decontamination of items.<br />

2. Place absorbent paper or linen towel in the bottom<br />

of trays before use.<br />

3. Unwrapped goods should only be processed on<br />

trays. Do not place items on the rack or the bottom<br />

of the sterilizer.<br />

4. Liquids must not exceed 75% of capacity of flask.<br />

Use self-venting caps.<br />

5. Open all hinged instruments.