MAINTENANCE<br />
50<br />
Warning<br />
Pressurized steam will be discharged<br />
when the relief valve is<br />
opened. To prevent burns, place a<br />
steam barrier (such as a rolled<br />
towel) around the opening in the<br />
back of the unit. To prevent burns,<br />
wear gloves or use an extension<br />
device if it becomes necessary to<br />
operate the pull ring.<br />
Note<br />
In the event that you have lost the<br />
supplied drain tubing, you can<br />
drain the reservoir as follows:<br />
1. Tilt the front of the sterilizer up<br />
until you can access the small<br />
panel on the bottom front of the<br />
sterilizer, beneath the drain<br />
assembly.<br />
2. Remove the screws securing the<br />
panel. Remove the panel.<br />
3. Have a suitable container prepared<br />
for the water in the reservoir.<br />
Disconnect the reservoir drain<br />
tube from the quick disconnect fixture<br />
on the front of the sterilizer.<br />
The reservoir will drain.<br />
Note<br />
Do not add more than 7 liters of<br />
water to the reservoir after cleaning.<br />
Monthly Maintenance<br />
Pressure Relief Valve Check<br />
To ensure that the pressure relief valve is functioning properly,<br />
you must have a qualified person check the valve each<br />
month.<br />
The pressure relief valve is located on the pressure<br />
line in the back of the unit and is accessed through a round<br />
access hole in the back panel. A suitable tool (screwdriver,<br />
rod, etc.) will be needed to test the pressure valve.<br />
1. Run a sterilize cycle at 135°C (275°F) without a<br />
load in the chamber.<br />
2. Allow the pressure in the chamber to reach 30 psi<br />
(206 kPa).<br />
3. Insert a suitable tool through the access hole and<br />
push the lever on the pressure relief valve. Push<br />
the valve lever away from the valve to open the<br />
valve. Hold the valve open for 3 seconds. <strong>Steam</strong><br />
should discharge freely.<br />
4. Release the valve lever. The steam discharge<br />
should cease entirely.<br />
5. After checking the valve, turn the sterilizer off by<br />
pressing "Stop" to prevent overheating.<br />
If the valve malfunctions, the valve must be replaced.<br />
Cleaning Reservoir<br />
1. The reservoir drain assembly is located on the<br />
lower right front side of the sterilizer cabinet. Place<br />
a suitable container below the drain assembly.<br />
Connect the drain adapter and tube to the drain<br />
assembly and drain the reservoir into the container.<br />
2. When the reservoir is empty, wipe out the inside of<br />
the reservoir with a clean, lint-free cloth. Take care<br />
not to push debris back into the water line.<br />
3. Refill the reservoir with distilled or low quality<br />
deionized water (minimum 0.5 megohms/cm, maximum<br />
1.5 megohms/cm) and replace the reservoir<br />
cover.